CASE REPORT
Year : 2011  |  Volume : 1  |  Issue : 2  |  Page : 161-163

The stunned atrial lead: Transient malfunction of a permanent atrial pacer lead following acute myocardial infarction


1 Department of Pediatric Cardiology, Medical University of South Carolina, MUSC, 601, Children's Hospital, 165, Ashley Avenue, Charleston, SC - 29425, USA
2 Department of Cardiology, Yale University, Bridgeport Hospital, 267 Grant Street, Bridgeport, CT - 06460, USA

Date of Web Publication12-Sep-2011

Correspondence Address:
Shailendra Upadhyay
Department of Pediatric Cardiology, Medical University of South Carolina, MUSC, 601, Children's Hospital, 165, Ashley Avenue, Charleston, SC - 29425
USA
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2229-5151.84806

Rights and Permissions
   Abstract 

Proximal right coronary artery occlusion caused transient loss of sensing and capture of the atrial lead of a permanent dual-chamber pacemaker. Forty-five days after percutaneous revascularization, the atrial lead was discovered to be functioning normally. We hypothesize that ischemia of the right atrium caused stunning of the atrial myocardium at the pacer-lead interface, which gradually improved following percutaneous coronary intervention (PCI), leading to return of lead function over time. So far only one similar case has been described in the literature.

Keywords: Atrial lead, myocardial infarction, myocardial stunning, pacemaker


How to cite this article:
Upadhyay S, Marshalko S, McPherson C. The stunned atrial lead: Transient malfunction of a permanent atrial pacer lead following acute myocardial infarction. Int J Crit Illn Inj Sci 2011;1:161-3

How to cite this URL:
Upadhyay S, Marshalko S, McPherson C. The stunned atrial lead: Transient malfunction of a permanent atrial pacer lead following acute myocardial infarction. Int J Crit Illn Inj Sci [serial online] 2011 [cited 2023 Mar 22];1:161-3. Available from: https://www.ijciis.org/text.asp?2011/1/2/161/84806

Malfunction of atrial lead of a permanent transvenous pacemaker following acute myocardial is rare. Here we describe this unusual case of transient atrial lead malfunction of a dual chamber trans-venous permanent pacemaker following acute inferior wall myocardial infarction.


   Case Report Top


A 78-year-old average-built man with previous history of coronary artery disease was hospitalized with chest pain of few hours duration. His vital signs and physical examination were unremarkable. Two years earlier, he had presented with syncope. Thorough investigation had revealed carotid hypersensitivity (3 second pause after right carotid massage) and mixed cardioinhibitory and vasodepressor response to upright tilt. A dual-chamber pacemaker (Guidant Discovery II DR) was implanted. The atrial lead was Guidant (4480-318341) passive-fixation bipolar electrode and the ventricular lead was Guidant (4457-39232665) passive-fixation bipolar electrode. He did well thereafter, experiencing no further syncope. Pacer checks every 6 months demonstrated normal function of both leads. Pacemaker interrogation 26 days prior to the present admission showed normal function of the atrial and ventricular leads. On the day of admission, the electrocardiogram demonstrated sinus rhythm, ST-segment elevation in the inferior leads, ST-segment depression in lead V2, normal AV conduction, and no PR-segment depression. Pacer stimuli occurred at 65 bpm, uninhibited by the prevailing sinus rhythm [Figure 1].
Figure 1: Admission ECG. The tracing demonstrates sinus rhythm at 68 bpm, normal conduction, inferior current of injury and ST-Segment depression in leads I, AVL, and V2. Atrial pacer stimuli occur at 65 pulse per minute, uninhibited by the prevailing sinus rhythm, indicating lead sensing failure. Atrial pacing at maximum output (7.5 V, 1.9 ms) demonstrated no atrial capture (data not shown).

Click here to view


Emergent cardiac catheterization demonstrated total occlusion of the proximal right coronary artery. Angioplasty was performed at the take-off on an atrial branch. Normal flow was restored and a stent was placed. A distal lesion was opened by balloon angioplasty [Figure 2]a and b. After catheterization, interrogation of the pacer demonstrated no measurable P-wave: at sensitivity of 0.15 mV the atrial lead detected only a far-field QRS. It failed to capture the atrium at the highest programmable output (7.0 V at 1.9 ms). The ventricular lead functioned normally. The basic metabolic panel and electrolytes were normal. Chest radiograph demonstrated stable atrial lead position with no visible discontinuities. The device was programmed to the rate-adaptive, ventricular inhibited (VVIR) mode. Two days later, the patient was clinically stable and ready for discharge. Pacer interrogation was unchanged; the atrial lead function had not improved. Forty-five days later, he returned to the Pacer Clinic complaining of effort intolerance and fatigue. Interrogation revealed normal atrial sensing (P-wave = 3.0 mV) and capture (threshold = 0.8 V at 0.4 ms). The pacer was reprogrammed to the dual-chamber mode. Within 48 hours, his symptoms improved. Follow-up pacer interrogations 2 months and 8 months later demonstrated stable function of the atrial lead [Table 1].
Table 1: Summary of atrial and ventricular lead testing before and after the day of acute inferior wall myocardial infarction (29th June, 2004)

Click here to view
Figure 2: Right coronary angiogram in the LAO cranial projection before (a) and after (b) angioplasty of the proximal occlusion. The atrial (A) and ventricular (B) leads are seen. TIMI grade 0 fl ow before angioplasty was restored to TIMI grade 3. A large right atrial branch (black arrow) is seen originating at the point of the vessel occlusion.

Click here to view



   Discussion Top


This case demonstrates transient malfunction of a permanent passive-fixation atrial pacer lead at the time of acute myocardial infarction. The infarction was due to total occlusion of the proximal right coronary artery at the take-off of the atrial branch. Since malfunction of the lead preceded angioplasty, the stent placement in the right coronary artery cannot be implicated in the lead malfunction. We conclude that transient ischemia of the right atrial myocardium at the pacer-lead interface caused transient loss of sensing and pacing functions. This persisted for at least 48 hours but resolved within 2 months. The most plausible hypothesis to explain these observations is stunning of the atrial myocardium due to atrial ischemia. Neither PR-segment depression nor atrial fibrillation - two clinical phenomena that often indicate atrial infarction - were observed in this case. [1] Chiliang et al. have reported one similar case. In that study, the atrial lead malfunction was found to have normalized 40 days after stent implantation of right coronary artery. [2]


   Conclusion Top


This case demonstrates that transient malfunction of a permanent atrial pacer lead may result from occlusion of the right coronary artery proximal to the atrial branch. In the absence of other electrocardiographic features of atrial infarction, transient malfunction of an atrial pacing lead should be included as a criterion for atrial infarction in patients with pacemakers. In this clinical setting, a new atrial lead should not be urgently implanted; the patient should be followed with the expectation that lead function may normalize over time. The precise time course of such recovery is not fully known. Thus, should other physicians encounter similar patients, it would be informative to follow patients at close intervals to determine how quickly the function may normalize.

 
   References Top

1.Sivertssen E, Hoel B, Bay G, Jörgensen L. Electrocardiographic atrial complex and acute atrial myocardial infarction. Am J Cardiol 1973;31:450-6.  Back to cited text no. 1
    
2.Chen C, Fang CC, Wang SP. Reversible atrial lead dysfunction of DDD pacemaker after recent inferior myocardial infarction-a case report. Angiology 2004;55:451-4.  Back to cited text no. 2
[PUBMED]  [FULLTEXT]  


    Figures

  [Figure 1], [Figure 2]
 
 
    Tables

  [Table 1]


This article has been cited by
1 Loss of Right Atrial Pacing Lead Capture Due to Myocardial Infarction Obscuring Diagnosis on Electrocardiogram
James A. McCaffrey, Bishnu P. Dhakal, Ashwin S. Nathan, Juan Ortega-Legaspi, Paul N. Fiorilli, Paul J. Mather, Gregory E. Supple
JACC: Case Reports. 2022; 4(14): 890
[Pubmed] | [DOI]
2 Case report: Pacemaker lost capture after acute myocardial infarction in a patient with left circumflex coronary artery occlusion
Zhihong Wu, Jianjun Tang, Qingyi Zhu, Lin Hu, Zhenjiang Liu, Xuping Li, Qiming Liu, Shenghua Zhou, Mingxian Chen
Frontiers in Cardiovascular Medicine. 2022; 9
[Pubmed] | [DOI]
3 Transient rise in His-lead threshold due to acute myocardial infarction
Robert Hall,Daniel A Jones,Amal Muthumala,Roshan Weerackody,Afzal Sohaib,Christopher Monkhouse
Pacing and Clinical Electrophysiology. 2019;
[Pubmed] | [DOI]
4 Acute Inferior Wall ST Segment Elevation Myocardial Infarction Caused Atrial Lead Dysfunction Reversed by Primary Percutaneous Coronary Intervention
Hui-Ting Wang,Yung-Lung Chen,Chiung-Jen Wu
International Heart Journal. 2018;
[Pubmed] | [DOI]
5 Pacemaker malfunction after acute myocardial infarction in a patient with wrap-around left anterior descending artery supplying the right ventricular apex
Tomohiko C. Umei,Toru Awaya,Osamu Okazaki,Hisao Hara,Yukio Hiroi
Journal of Cardiology Cases. 2018;
[Pubmed] | [DOI]
6 Long-term performance of right ventricular pacing leads: risk factors associated with permanent right ventricular pacing threshold increase
Hui Peng,Zhijun Sun,Heping Zhang,Wenying Ma
Journal of Interventional Cardiac Electrophysiology. 2018;
[Pubmed] | [DOI]
7 Early improvement of pacing threshold following primary right coronary angioplasty
Fernando Pivatto Júnior,Diego Chemello,Geris Mazzutti,Maurício Pimentel,Paola Rabaioli,Leandro Zimerman
HeartRhythm Case Reports. 2016;
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Case Report
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed5429    
    Printed213    
    Emailed0    
    PDF Downloaded82    
    Comments [Add]    
    Cited by others 7    

Recommend this journal