ORIGINAL ARTICLE |
|
Year : 2013 | Volume
: 3
| Issue : 4 | Page : 262-268 |
|
Evaluation of tracheal cuff pressure variation in spontaneously breathing patients
Gustavo A Plotnikow1, Nicolas Roux2, Viviana Feld2, Emiliano Gogniat2, Dario Villalba2, Noelia Vairo Ribero2, Marisa Sartore2, Mauro Bosso2, Corina Quiroga1, Valeria Leiva1, Mariana Scrigna2, Facundo Puchulu2, Eduardo Distéfano2, Jose Luis Scapellato3, Dante Intile1, Fernando Planells2, Diego Noval2, Pablo Buñirigo2, Ricardo Jofré2, Ernesto Díaz Nielsen2
1 Respiratory Care Services, Clínica Basilea; Intensive Care Unit, Sanatorio Anchorena, Buenos Aires, Argentina 2 Respiratory Care Services, Clínica Basilea, Argentina 3 Intensive Care Unit, Sanatorio Anchorena, Buenos Aires, Argentina
Date of Web Publication | 2-Jan-2014 |
Correspondence Address: Gustavo A Plotnikow Gral, Pueyrredón 261, 1888, Florencio Varela, Buenos Aires Argentina
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/2229-5151.124148
Abstract | | |
Background: Most of the studies referring cuff tubes' issues were conducted on intubated patients. Not much is known about the cuff pressure performance in chronically tracheostomized patients disconnected from mechanical ventilation. Objective : To evaluate cuff pressure (CP) variation in tracheostomized, spontaneously breathing patients in a weaning rehabilitation center. Materials and Methods: Experimental setup to test instruments in vitro, in which the gauge (TRACOE) performance at different pressure levels was evaluated in six tracheostomy tubes, and a clinical setupin which CP variation over 24 h, every 4 h, and for 6 days was measured in 35 chronically tracheostomized clinically stable, patients who had been disconnected from mechanical ventilation for at least 72 h. The following data were recorded: Tube brand, type, and size; date of the tube placed; the patient's body position; the position of the head; axillary temperature; pulse and respiration rates; blood pressure; and pulse oximetry. Results : In vitro difference between the initial pressure (IP) and measured pressure (MP) was statistically significant (P < 0.05). The difference between the IP and MP was significant when selecting for various tube brands (P < 0.05). In the clinical set-up, 207 measurements were performed and the CP was >30 cmH 2 O in 6.28% of the recordings, 20-30 cmH 2 O in 42.0% of the recordings, and <20 cmH 2 O in 51.69% of the recordings. Conclusion : The systematic CP measurement in chronically tracheostomized, spontaneously breathing patients showed high variability, which was independent of tube brand, size, type, or time of placement. Consequently, measurements should be made more frequently. Keywords: Artificial airway, cuff pressure, decannulation, prolonged mechanical ventilation, tracheostomy, ventilator-associated pneumonia
How to cite this article: Plotnikow GA, Roux N, Feld V, Gogniat E, Villalba D, Ribero NV, Sartore M, Bosso M, Quiroga C, Leiva V, Scrigna M, Puchulu F, Distéfano E, Scapellato JL, Intile D, Planells F, Noval D, Buñirigo P, Jofré R, Nielsen ED. Evaluation of tracheal cuff pressure variation in spontaneously breathing patients. Int J Crit Illn Inj Sci 2013;3:262-8 |
How to cite this URL: Plotnikow GA, Roux N, Feld V, Gogniat E, Villalba D, Ribero NV, Sartore M, Bosso M, Quiroga C, Leiva V, Scrigna M, Puchulu F, Distéfano E, Scapellato JL, Intile D, Planells F, Noval D, Buñirigo P, Jofré R, Nielsen ED. Evaluation of tracheal cuff pressure variation in spontaneously breathing patients. Int J Crit Illn Inj Sci [serial online] 2013 [cited 2022 Aug 16];3:262-8. Available from: https://www.ijciis.org/text.asp?2013/3/4/262/124148 |
Introduction | |  |
Microaspiration of oropharyngeal contents caused by underinflation of the tracheal cuff in intubated patients can lead to ventilator-associated pneumonia (VAP) causing morbidity and mortality in intubated, mechanically ventilated, and tracheostomized spontaneously breathing patients. It is recommended that cuff pressure (CP) levels be kept at 20-30 cmH 2 O. [1],[2],[3],[4],[5] When the CP falls to <20 cmH 2 O, the risk of VAP increases four-fold; in comparison, at >30 cmH 2 O, the cuff may damage the tracheal tissue. [4],[5],[6],[7],[8],[9],[10] Despite the control of CP, underinflation and overinflation occur frequently in the intensive care unit (ICU). Most guidelines regarding the control and regulation of CP are based on surveys and they vary considerably, suggesting checks every 2-12 h. [11],[12],[13],[14]
Several factors influence CP, including changes in tracheal muscle tone, body temperature, anesthetic gas diffusion inside the cuff, endotracheal tube (ETT) position, respiratory system impedance, level of consciousness, and the position of the body, head, and neck. [15],[16],[17],[18] Although continuous CP control monitors variation throughout the day in mechanically ventilated patients with an ETT, allowing for pressure regulation, the incidence of VAP does not decrease. [6],[19],[20],[21] There is no agreement on the best CP monitoring method; however, the most commonly used method is direct control with a pressure gauge. [22],[23],[24],[25],[26],[27]
Most studies of variation and regulation of CP have been performed using intubated and mechanically ventilated patients, but studies on CP in tracheostomized patients are scarce. Our aim was to evaluate CP variation in tracheostomized, spontaneously breathing patients in a weaning rehabilitation center.
Materials and Methods | |  |
This two-part study was carried out at a weaning center in Buenos Aires, Argentina, from March 1 to May 31, 2010. It consisted of an experimental setup to test instruments in vitro and a clinical set-up, in which CP variation was measured in tracheostomized patients.
To monitor the pressure, we used a TRACOE Cuff Pressure Monitor (Medical, Frankfurt, Germany), Rüsch Safety Clear tracheostomy tubes (numbers 8, 9, and 10); a Portex Blue Line Ultra Suctionaid Tracheostomy Tube Kit (number 8); Mallinckrodt Tracheosoft Evac ETT (numbers 8 and 10); and a 10 cm × 28 mm rigid plastic cylinder.
Experimental setup
Using the in vitro setup, we evaluated the gauge performance at different pressure levels in six tracheostomy tubes (APPENDIX 1).
A plastic cylinder was used to simulate the trachea; a tracheostomy tube was inserted and the cuff was inflated to the initial pressure (IP) level with a TRACOE pressure gauge. The gauge was removed, reattached, and measured pressure (MP) was recorded in an Excel spreadsheet. If necessary, the pressure was corrected to the initial value. This sequence was repeated ten times, with different IPs each time, starting at 40 cmH 2 O and decreasing by 2 cmH 2 O for each measurement to reach 10 cmH 2 O on the last measurement. This process was repeated for all six tracheostomy tubes.
Clinical setup
The aim of the first part of this study was to evaluate CP variations over 24 h in 35 patients aged >18 years, who were chronically tracheostomized, clinically stable, and who had been disconnected from mechanical ventilation for at least 72 h. The patients were excluded if they required tube replacement, mechanical ventilation, transfer to another center, or discharge during the study period. The tube brands used were the Mallinckrodt Tracheosoft Evac in 20 patients; Portex Blue Line Ultra Suctionaid in 3 patients; and Rüsch Safety Clear in 12 patients. The patients were randomly allocated to three groups. Each patient was then randomly assigned to a preset measurement scheme for 6 days [Table 1]. All assignments were made via sealed envelopes. According to this scheme, all patients had their CP calibrated to 30 cmH 2 O (IP) at 8:00 AM. Then, for the following 6 days, each patient's CP was measured and calibrated only once every day at a pre-set time, which differed for each patient daily. In this way, the CP was monitored over 24 h, every 4 h, and for 6 days. This scheme was designed by our team to avoid lapses in the recorded data and pressure loss resulting from consecutive measurements.
The following data were recorded using forms and an Excel spreadsheet: Tracheostomy tube brand, type, and size; date the tube was placed; the patient's body position (dorsal decubitus, lateral decubitus, or seated); the position of the head (neutral or rotated); axillary temperature; pulse and respiration rates; blood pressure; and pulse oximetry [Table 2].
Categorical variables are presented as frequencies (%). Continuous variables are described as the mean ± SD. Normality was tested using Shapiro-Wilks and Kolmogorov-Smirnov testsAnalysis of variance (ANOVA) for repeated measurements and multiple linear regression were used to analyze the data. P ≤ 0.05 was considered significant. SPSS v. 19 (IBM, Armonk, NY, USA) was used for all statistical analyses.
Results | |  |
Experimental setup
A total of measurements were performed in vitro [Table 3]. The difference between the IP and MP was statistically significant (P < 0.05). Moreover, the difference between the IP and MP was significant when selecting for various tube brands (P < 0.05). The Table in APPENDIX 1 details our laboratory measurements. | Table 3: In vitro cuff pressure measurements (experimental setup: 960 measurements)
Click here to view |
Clinical setup
A total of 207 measurements were performed. The CP was >30 cmH 2 O in 6.28% of the recordings, 20-30 cmH 2 O in 42.0% of the recordings, and <20 cmH 2 O in 51.69% of the recordings. [Figure 1] shows significant variation in the CP. The mean time between intubation and the first CP measurement was 69.9 days (range, 7-209 days). The time between tube placement and first measurement, which was divided into three categories (0-2, 2-4, and 4-6 months), showed a mean CP of 24.22, 18.37, and 17.85 cmH 2 O, respectively. | Figure 1: CP variation in 1 day. CI (confidential interval) 95% P > 0.05
Click here to view |
A multiple linear regression model showed no significant difference in CP depending on the tube brand, size, and time of placement. The variation in CP for the various tube brands is shown in [Figure 2]. No significant difference was found in CP for the different body and head positions: Dorsal decubitus/neutral head was at 21.87 (±11.44) cmH 2 O; dorsal decubitus/rotated head was at 20.23 (±9.52) cmH 2 O; lateral decubitus/neutral head was at 20.03 (±6.44) cmH 2 O; lateral decubitus/rotated head was at 17.45 (±5.53) cmH 2 O; seated/neutral head was at 17.6 (±4.82) cmH 2 O; and seated/rotated head was at 19.25 (±8.54) cmH 2 O. The CP tended to be lower in the lateral decubitus/rotated head and seated/neutral head positions. | Figure 2: CP variation over 1 day for each of the tracheostomy tube brands. CI 95% P > 0.05
Click here to view |
Discussion | |  |
We assessed the variation in CP in spontaneously breathing, chronically tracheostomized patients for 24 h and then the CP variation when it was controlled every 4 h, and we suggest that measurements should be taken at least every 4 h.
In the experiment, we found that the mean CP decrease for each tracheostomy tube brand was very similar at every IP level. This might be explained by the characteristic pressure loss at the time of measurement in each tube, and the losses might be caused by leaks in the cuff pilot valve when the pressure gauge was inserted or removed; however, they were not caused by the compressible volume of the equipment. This concurs with Blanch et al., who found that the compressible volume of four brands of pressure gauge was not significant.
Although we found no significant differences in the pressures according to the age of the tube, the pressure loss tended to be greater in older tubes. A larger sample might have allowed us to confirm this tendency. Clinically, Portex tubes had the most recordings >30 cmH 2 O [Figure 2]. These tubes showed stable performance in terms of pressure loss in the experiment.
Controlling the CP every 4 h in the same tube for 24 h would have led to large pressure loss within the period studied, affecting the true determination of CP variation. Using a correction table in the experiment, we noted that the pressure loss resulting from consecutive measurements produced values below the minimum recommended values.
The continuous measurement technique is the most reliable technique for describing the CP variation. [19],[20] However, it is very difficult to apply this method in clinical practice. Published studies using continuous measurement did not assess the CP variation over 24 h, indicating that the CP variation was not controlled during the night. On analyzing the data, we found that the CP variation at night was similar to that recorded during the day. The studies based on autoregulation systems do not describe the CP variation, but they manage to keep it constant. [6],[20]
We found that 51.69% of the CP measurements were <20 cmH 2 O, while 6.28% were >30 cmH 2 O, and 42.52% were normal, within the range of 20-30 cmH 2 O. In a study of continuous measurement in ETT patients ventilated mechanically, Sole et al., reported that, in 54% of the measurements, the CP varied from 20 to 30 cmH 2 O, while it was >30 cmH 2 O 16% of the time and <20 cmH 2 O 30% of the time. [21] They reported a mean 24.6 cmH 2 O for 12 h of continuous measurement, while our mean CP was 19.75 ± 9.08 cmH 2 O over 24 h.
Unlike the study by de Godoy et al., we did not find significant CP changes with body position. [17] We described the patient's position at the time of measurement, while de Godoy assessed the CP after a change in the patient's position, which might have affected the CP. We did not observe any influence of the position of the head and neck at the time of measurement on the CP. Inoue et al., described the pressure changes in patients with head extension or flexion, and they associated the variation with ETT movement. [18] Although we found no significant difference in our population, the CP tended to be lower in the lateral decubitus or seated position versus dorsal decubitus.
Inoue et al., and de Godoy et al., analyzed the CP in relation to the ETT. The ETT length and the traction of the head/neck movement on the ETT might have more influence on the ETT CP than on a tracheostomy tube cuff, probably because the ETT parallels the cervical spine, while tracheotomy tubes are inserted at lower cervical level.
The limitations of our study is that it was not a randomized control trial, as that would have allowed us to compare the pressure control method and the standard control every 8 h. Another serious limitation was that the tubes used in the clinical setup were those that the patients already had when we started the study; brand new tubes in all patients may have shown different results.
Conclusion | |  |
The systematic CP measurement in chronically tracheostomized, spontaneously breathing patients showed high variability, independent of tube brand, size, type, or time of placement. Therefore, values above or below the recommended values may be found at any measurement time, even after daily calibration. Consequently, measurements should be made more frequently. Nevertheless, measurements taken every 4 h are associated with a progressive reduction in CP, which might be associated with micro-aspiration. Therefore, we suggest checking the CP daily, or at least every 4 h, and maintaining the CP at the upper limit (30 cmH 2 O) to avoid CP decrements below the lower limit. Patients hospitalized in rehabilitation centers are generally active and perform activities out of bed. To our knowledge, this is the first study of tube performance and CP variation in such a population. Further research is needed to assess the clinical impact CP variation during the measurement process.

References | |  |
1. | American Thoracic Society; Infectious Diseases Society of America. Guidelines for the management of adults with hospital-acquired, ventilator-associated, and healthcare-associated pneumonia. Am J Respir Crit Care Med 2005;171:388-416.  |
2. | Bernhard WN, Yost L, Joynes D, Cothalis S, Turndorf H. Intracuff pressures in endotracheal and tracheostomy tubes. Related cuff physical characteristics. Chest 1985;87:720-5.  |
3. | Diaz E, Rodríguez AH, Rello J. Ventilator-associated pneumonia: Issues Related to the artificial airway. Respir Care 2005;50:900-6.  |
4. | Rello J, Soñora R, Jubert P, Artigas A, Rué M, Vallés J. Pneumonia in intubated patients: Role of respiratory airway care. Am J Respir Crit Care Med 1996;154:111-5.  |
5. | Kollef MH. The prevention of ventilator-associated pneumonia. N Engl J Med 1999;340:627-34.  |
6. | Valencia M, Ferrer M, Farre R, Navajas D, Badia JR, Nicolas JM, et alAutomatic control of tracheal tube cuff pressure in ventilated patients in semirecumbent position: A randomized trial. Crit Care Med 2007;35:1543-9.  |
7. | Bishop MJ. Mechanisms of laryngotracheal injury following prolonged tracheal intubation. Chest 1989;101:185-6.  |
8. | Vyas D, Inweregbu K, Pittard A. Measurement of tracheal tube cuff pressure in critical care. Anaesthesia 2002;57:275-7.  |
9. | Braz JR, Navarro LH, Takata IH, Nascimento Júnior P. Endotracheal tube cuff pressure: Need for precise measurement. Sao Paulo Med J 1999;117:243-7.  |
10. | Blanch PB. Laboratory evaluation of 4 brands of endotracheal tube cuff inflator. Respir Care 2004;49:166-73.  |
11. | Tyler DO, Clark AP, Ogburn-Russell L. Developing a standard for endotracheal tube cuff care. Dimens Crit Care Nurs 1991;10:54-61.  |
12. | Crimlisk JT, Horn MH, Wilson DJ, Marino B. Artificial airways: A survey of cuff management practices. Heart Lung 1996;25:225-35.  |
13. | Sole ML, Byers JF, Ludy JE, Ostrow CL. Suctioning techniques and airway management practices: Pilot study and instrument evaluation. Am J Crit Care 2002;11:363-8.  |
14. | Sole ML, Byers JF, Ludy JE, Zhang Y, Banta CM, Brummel K. A multisite survey of suctioning techniques and airway management practices. Am J Crit Care 2003;12:220-30.  |
15. | Mehta S, Mickiewicz M. Pressure in large volume, low pressure cuffs: Its significance, measurement and regulation. Intensive Care Med 1985;11:267-72.  |
16. | Torres A, El-Ebiary M, Soler N, Montón C, Fàbregas N, Hernández C. Stomach as a source of colonization of the respiratory tract during mechanical ventilation: Association with ventilator-associated pneumonia. Eur Respir J 1996;9:1729-35.  |
17. | Godoy AC, Vieira RJ, Capitani EM. Endotracheal tube cuff pressure alteration after changes in position in patients under mechanical ventilation. J Bras Pneumol 2008;34:294-7.  |
18. | Inoue S, Takauchi Y, Kuro M, Ninaga H. Effects of changes in head and neck position on a tracheal tube cuff. Masui 1998;47:1069-72.  |
19. | Sole ML, Aragon D, Bennett M, Johnson RL. Continuous measurement of endotracheal tube cuff pressure: How difficult can it be? ACCN Adv Crit Care 2008;19:235-43.  |
20. | Farré R, Rotger M, Ferre M, Torres A, Navajas D. Automatic regulation of the cuff pressure in endotracheally intubated patients. Eur Respir J 2002;20:1010-3.  |
21. | Sole ML, Penoyer DA, Su X, Jimenez E, Kalita SJ, Poalillo E, et al. Assessment of endotracheal cuff pressure by continuous monitoring: A pilot study. Am J Crit Care 2009;18:133-43.  |
22. | O′Donnell JH. Endotracheal tube intra-cuff pressure initially and during anesthesia including nitrous oxide. CRNA 1995;6:79-85.  |
23. | Hess DR. Managing the artificial airway. Respir Care 1999;44:759-71.  |
24. | Fernandez R, Blanch L, Mancebo J, Bonsoms N, Artigas A. Endotracheal tube cuff pressure assessment: Pitfalls of finger estimation and need for objective measurement. Crit Care Med 1990;18:1423-6.  |
25. | Stewart SL, Secrest JA, Norwood BR, Zachary R. A comparison of endotracheal tube cuff pressures using estimation techniques and direct intracuff measurement. AANA J 2003;71:443-7.  |
26. | Sengupta P, Sessler DI, Maglinger P, Wells S, Vogt A, Durrani J, et al. Endotracheal tube cuff pressure in three hospitals, and the volumen required to produce an appropriate cuff pressure. BMC Anesthesiol 2004;4:8.  |
27. | Mallampati SR. Airway management. In: Barash PG, Cullen BF, Stoelting RK, editors. Clinical Anesthesia. Philadelphia: Lippincott-Raven; 1997. p. 583-5.  |
[Figure 1], [Figure 2]
[Table 1], [Table 2], [Table 3]
This article has been cited by | 1 |
Continuous Endotracheal Tube Cuff Pressure Control Decreases Incidence of Ventilator-Associated Pneumonia in Patients with Traumatic Brain Injury |
|
| Mehmet Salih Sevdi,Serdar Demirgan,Kerem Erkalp,Onat Akyol,Funda Gumus Ozcan,Hasan Cem Guneyli,Mehmet Can Tunali,Aysin Selcan | | Journal of Investigative Surgery. 2021; : 1 | | [Pubmed] | [DOI] | | 2 |
Slaugytoju žinios apie tracheostomos vamzdelio manžetes priežiura prieš ir po mokymo |
|
| Silva Kostylioviene,Alina Vaškelyte,Dovile Grinkeviciute | | Sveikatos mokslai. 2017; 26(6): 213 | | [Pubmed] | [DOI] | | 3 |
Comparison of endotracheal tube cuff pressure values before and after training seminar |
|
| Ayça Tuba Dumanli Özcan,Cihan Döger,Abdülkadir But,Isik Kutlu,Semsi Mustafa Aksoy | | Journal of Clinical Monitoring and Computing. 2017; | | [Pubmed] | [DOI] | |
|
 |
 |
|