ORIGINAL ARTICLE
Year : 2017  |  Volume : 7  |  Issue : 3  |  Page : 156-162

Comparison of pulmonary artery catheter, echocardiography, and arterial waveform analysis monitoring in predicting the hemodynamic state during and after cardiac surgery


1 Intensive Care Unit, University Hospital Geelong, Barwon Health, Geelong, Victoria, Australia
2 Intensive Care Unit, University Hospital Geelong, Barwon Health, Geelong, Victoria; Deakin University School of Medicine, Barwon Health, Geelong, Victoria, Australia
3 Department of Cardiothoracic Surgery, University Hospital Geelong, Barwon Health, Geelong, Victoria; Department of Epidemiology and Preventive Medicine, Monash University, Melbourne, Victoria, Australia
4 Department of Anaesthesia, University Hospital Geelong, Barwon Health, Geelong, Victoria, Australia
5 Department of Epidemiology and Preventive Medicine, Australian and New Zealand Intensive Care Research Centre, Monash University, Melbourne, Victoria, Australia

Date of Web Publication12-Sep-2017

Correspondence Address:
Paul Power
Intensive Care Unit, University Hospital Geelong, P.O. Box 281, Geelong, Victoria, 3220
Australia
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2229-5151.214411

Rights and Permissions
   Abstract 

Objective: The aim of this trial was to determine whether Flotrac Vigileo™ (FV™) provides a reliable representation of the hemodynamic state of a cardiac surgical patient population when compared to pulmonary artery catheter (PAC) and echocardiography in the peril-operative period.
Design: This was a prospective observational trial comparing perioperative hemodynamic states using transesophageal echocardiography (TEE), transthoracic echocardiography (TTE), FV™ and PAC during and post cardiothoracic surgery.
Setting: Tertiary regional hospital Intensive Care Unit (ICU).
Participants: 50 consecutive adult cardiothoracic patients with written consent provided.
Intervention: Comparison of the perioperative hemodynamic states using echocardiography, FV™ and PAC was performed. Evaluation of the hemodynamic state (HDS) was performed using TEE, TTE, PAC and FV™ during and after cardiac surgery. Data were compared between the three hemodynamic assessment modalities.
Main Outcome Measure: Predicted hemodynamic state.
Results: FV™ and PAC were shown to correlate poorly with TEE/TTE assessment of the hemodynamic state. Both PAC and FV™ showed significant discordance with echocardiographic assessment of the hemodynamic state.
Conclusions: In this trial, FV™ and PAC were shown to agree poorly with TTE/TEE assessment of the HDS in an adult cardiothoracic population. Agreement between the FV™ and PAC was also poor. Caution is recommended in interpreting isolated hemodynamic monitoring data. All hemodynamic monitoring devices have inherent sources of error. Caution is advised in interpreting any single device or measurement as a gold standard. We suggest that hemodynamic measuring devices such as FV™/PAC may act as triggers for a global hemodynamic assessment including consideration of TTE/TEE.

Keywords: Anesthesia, cardiaothoracic, hemodynamic monitoring, intensive care


How to cite this article:
Power P, Bone A, Simpson N, Yap CH, Gower S, Bailey M. Comparison of pulmonary artery catheter, echocardiography, and arterial waveform analysis monitoring in predicting the hemodynamic state during and after cardiac surgery. Int J Crit Illn Inj Sci 2017;7:156-62

How to cite this URL:
Power P, Bone A, Simpson N, Yap CH, Gower S, Bailey M. Comparison of pulmonary artery catheter, echocardiography, and arterial waveform analysis monitoring in predicting the hemodynamic state during and after cardiac surgery. Int J Crit Illn Inj Sci [serial online] 2017 [cited 2020 Nov 30];7:156-62. Available from: https://www.ijciis.org/text.asp?2017/7/3/156/214411


   Introduction Top


Invasive hemodynamic monitoring is commonly utilized in the perioperative care of cardiac surgery patients, and is an essential component of decision-making in relation to the hemodynamic state (HDS).[1] The pulmonary artery catheter (PAC) is used to measure cardiac output (CO) and systemic vascular resistance (SVR) to evaluate the hemodynamic status of a patient.[2] PACS have been in use for 30 years, despite equivocal evidence as to their utility in an intensive care environment.[3] PACs are associated with adverse events, including increased incidence of arrhythmias, providing false readings under certain conditions, and pulmonary artery rupture and death.[4] Noninvasive CO monitoring devices have some desirable properties but have been found to be variably reliable.[5],[6],[7],[8] Several recent trials have evaluated the efficacy of arterial waveform analysis monitoring using the FloTrac/Vigileo™ (FV™) (Edwards Lifesciences, Irvine, CA, USA) monitoring system.[9],[10],[11],[12],[13] FV™ data are extrapolated from the arterial pressure waveform after correction for body height and weight, age, and gender. Several iterations of FV™ software have made the device increasingly reliable.[10] The majority of trials have sought to validate the measurements of the FV™ compared to a gold standard (e.g., PAC, transthoracic echocardiography [TTE]). To safely manage patients during and after cardiac surgery, it is important to have reliable and regular assessments of the HDS.[14] The aim of this study was to determine whether FV™ provides a reliable representation of the patients' HDS when compared to PAC and echocardiography in the intra- and post-operative period.


   Methods Top


Patients and setting

University Hospital Geelong is a tertiary regional hospital with a level 3 Intensive Care Unit (ICU), and performs over 400 cardiac surgeries each year. The current method of hemodynamic monitoring during and after surgery includes a PAC in combination with perioperative transesophageal echocardiography (TEE).

Methods

After the Local Human Research Ethics Committee approval and obtaining prior written informed consent, fifty adult patients scheduled for elective or semi-urgent coronary artery bypass grafting (CABG) and or valve replacement/repair with cardiopulmonary bypass were included in the trial. Exclusion criteria were the presence of atrial fibrillation at the time of consent, planned intra-aortic balloon pump placement, and severe valvular dysfunction. FV™ monitors were loaned by Edwards Lifesciences for the duration of the trial; however, consumables were funded independently. Edwards Lifesciences were not involved in trial design or the trial itself, including data analysis or write-up.

Routine intra- and post-operative management

Anesthetic and postoperative management followed institutional clinical guidelines for intra- and post-operative care. After induction and intubation, a PAC (7.5 F, Edwards, Irvine, CA, USA) was placed through the internal jugular vein and correct positioning confirmed by pulmonary artery pressure waveform analysis. The PAC was then connected to an Edwards Vigilance II™ monitor for continuous CO monitoring. FV™ 3rd generation software was utilized. After radial artery cannulation, the arterial pressure waveform was confirmed as accurate and the pressure monitoring set and FV™ monitor were connected. Patients' height and weight were entered into the Vigilance II™ and FV™ monitors. All pressure transducers were referenced to the mid-axillary line and zeroed before recording of hemodynamic parameters.

Measurement of transthoracic echocardiography/transesophageal echocardiography hemodynamic parameters

The initial two TEE studies were performed in the operating theater as the part of routine cardiac assessment during surgery. An initial TEE was performed prebypass (TEE 1) and a 2nd postbypass (TEE 2). At these time points central venous pressure (CVP), cardiac index (CI) and SVR index (SVRI) from the PAC and FV™ were documented by an anesthetist. Postoperative TTE/TEE was performed after transfer to the ICU, when the patient was hemodynamically stable and able to be positioned on their left side. At the time of the TTE/TEE, CVP, CI and SVRI from the PAC and FV™ were documented by a registered nurse. In addition, after admission to ICU, CI, and SVRI from the FV™ were documented as the part of the trial until the PAC was removed. CVP was used in preference to pulmonary artery wedge pressure (PAWP) as PAWP is potentially hazardous and not routinely recorded in this cardiothoracic center. Hemodynamic parameters were collected and documented by an anesthetist or registered nurse with a specialty postgraduate qualification. TTE/TEE was performed by cardiac anesthetists/intensivists holding a qualification in critical care echocardiography (PTEeXAM®/H.A.R.T.scan®/Postgraduate Diploma in Clinical Ultrasound [PGDipCU]). Echocardiographic views and measurements were based on standard practice and commonly measured criteria during rapid assessment scanning protocols. Echocardiograms were classified into one of four HDSs using an algorithm based on modified H.A.R.T.scan® criteria [Appendix 1] [Additional file 1].[15],[16] Criteria included measurements of ventricular volume, systolic function, and interatrial septal motion. Left ventricular (LV) function was based on the measurement of fractional shortening (FS) or area change while the filling state was assessed using the LV end-diastolic diameter or area (depending on TTE or TEE). A vasodilated state was denoted by the echocardiographic findings of a hyperdynamic state (FS >44%) in the absence of criteria for “empty.” Using proprietary instructions on monitoring interpretation, measurements from the PAC (including CVP)[17],[18],[19] and FV™ were used to determine one of these four hemodynamic states for each device. Where HDSs were overlapping, the cardiac state was deemed to be the primary state (e.g., primary systolic failure [PSF] vs. vasodilated). All TTE/TEE studies were reviewed independently for agreement by a cardiac anesthetist and intensivist holding echocardiography qualifications (PTEeXAM®/PGDipCU).

Blinding

Vigilance II™ and Vigileo™ monitor screens were obscured at the time the anesthetist/intensivist was performing the TTE/TEE. Hemodynamic parameters were documented at the time the TTE/TEE was performed or immediately after.

Data

Each patient was allocated a number and patient identifiers were removed. Baseline data was collected for each patient. This included age, gender, procedure performed, length of ICU stay, APACHE II score, and duration of mechanical ventilation. TTE/TEE and corresponding hemodynamic data for the time points described above were entered into the patient database.

Statistical analysis

Continuous normally distributed variables were presented as means (standard deviation), whereas nonnormally distributed variables were reported as medians (interquartile range). Comparison was performed between multiple devices (TTE/TEE/PAC/FV™) for 4 HDSs, with agreement reported using kappa concordance.


   Results Top


Demographics

[Table 1] demonstrates patient characteristics. The cohort was predominantly male (86%), with the most frequent procedure CABG (58%) then aortic valve replacement (24%) surgery. The mean age was 68.5 years.
Table 1: Patient demographic data (n = 50)

Click here to view


Hemodynamic states

[Table 2] describes the HDS indicated by echocardiography. The majority of patients were classified as having an empty or normal HDS at the first echocardiogram (TEE), with an increased number of abnormal HDSs occurring peri- and immediately post-operatively.
Table 2: Assessment of hemodynamic state using echocardiography, n (%)

Click here to view


Pulmonary artery catheter and transthoracic echocardiography/transesophageal echocardiography

The strength of the relationship between PAC HDSs and TTE/TEE assessment was poor [Table 3]. PAC empty states only correctly predicted TTE/TEE empty state in 17/33 (51.5%) measurements, while a normal TTE state correlated with a PAC normal state on 29/80 (36.3%) measurements. Where echocardiography was normal, the PAC suggested an empty state in 39/58 (67.2%) measurements. A normal PAC state coincided with an empty TTE/TEE state in 10/33 (30.3%) measurements. The kappa correlation between echocardiography and PAC HDSs was 0.06.
Table 3: Agreement between echocardiography and pulmonary artery catheter for predicting hemodynamic state assessed using kappa statistics

Click here to view


FloTrac/Vigileo™ and transthoracic echocardiography/transesophageal echocardiography

The strength of the relationship between FV™ and echocardiographic states was also poor [Table 4]. The overall kappa concordance was 0.14. FV™ predicted the normal HDS in 56/83 (67.5%) TTE/TEE measurements, but only predicted empty in 12/30 (40%) measurements and PSF in 2/19 (10.5%) measurements when compared to echocardiography.
Table 4: Agreement between echocardiography and FloTrac/Vigileo™ for predicting hemodynamic state assessed using kappa statistics

Click here to view


FloTrac/Vigileo™ and pulmonary artery catheter

There was little relationship between HDSs when comparing FV™ and PAC [Table 5]. In particular, in this group, there was little correlation in the state of PSF between devices.
Table 5: Agreement between pulmonary artery catheter and FloTrac/Vigileo™ for predicting hemodynamic state assessed using kappa statistics

Click here to view


Primary systolic failure

Both PAC and FV™ tended to over-report the HDS of PSF when compared to echocardiography [Table 6] and [Table 7]. When the HDS was determined as PSF by PAC this was confirmed by echocardiography in 2/19 (10.5%) measurements and in 3/22 (13.6%) measurements when determined by FV™.
Table 6: Agreement between echocardiography and FloTrac/Vigileo™ for predicting hemodynamic state when the pulmonary artery catheter cardiac index was below 2.2 L/min/m2 (primary systolic failure) assessed using kappa statistics

Click here to view
Table 7: Agreement between echocardiography and pulmonary artery catheter for predicting hemodynamic state when the FloTrac/Vigileo™ cardiac index was below 2.2 L/min/m2 (primary systolic failure) assessed using kappa statistics

Click here to view


Comparison and the “normal” state

[Appendix 2] [Additional file 2] describes the HDSs when compared between the 3 devices. PAC and FV™ were comparably more reliable in their correlation with PSF states as determined by echocardiography (PAC = 3/9, FV™ = 2/7); although, numbers in this group were small. Both FV™ and PAC tended to over-record abnormal states when compared to echocardiography. When compared with TTE, FV™ predicted “normal” states in 56/83 (67.4%) while for PAC the rate was 29/90 (32%). The PAC recorded CI <2.2 in 6/90 (6.7%) readings simultaneously with “normal” TTE's while FV™ recorded 1/90 (1.1%) of these instances. FV™ predicted the “normal” HDS in 56 out of 83 “normal” patients as determined by echocardiography (67%). PAC predicted the “normal” HDS in 29 of 90 “normal” patients as determined by echocardiography (32%). At T1, for the same patient cohort, 37 abnormal HDSs were suggested by PAC (vs. 25 for FV™ and 15 for TEE/TTE).


   Discussion Top


This trial was designed to compare the HDS based on information from FV™, echocardiography, and PAC in a cardiothoracic surgical population. While overall correlation was poor, our data suggest that both FV™ and PAC may be useful in acting as triggers for assessment of the HDS.

To date, literature on FV™ has reported variable results. FV™ has been extensively studied in a number of settings. Many of these studies involve earlier software iterations,[20],[21] or were used in discrete clinical contexts such as sepsis[12] or volume expansion.[22] There are few studies that are based in the intensive care setting, with a percentage error between compared devices of 33% and 40% suggesting that the use of FV™ in this setting may be less reliable.[23] More recent literature is mixed. A study of 40 cardiac surgery patients demonstrated percentage error rates of up to 61.5%[24] while an error rate of 66.5% was demonstrated when comparing FV™ with PAC bolus thermodilution in a cohort of 25 cardiac surgery patients.[10]

Recent research using FV™ has focused on absolute values rather than the overall HDS[15] and seldom using TTE/TEE for comparison.[13] The evidence for the accuracy of FV™ data in a variety of clinical settings is mixed, although the correlation with invasive monitors appears to be improving through software iterations[12],[21] Questions still remain over the accuracy of FV™ in the setting of tachyarrythmias and valvular heart disease;[15] although, FV™ has been compared more favorably with TTE derived CO measurements when these patients groups have been excluded from the study.[13]

Sources of error from previous studies

The accuracy of FV™ in measuring CO under certain conditions has been questioned. Suehiro et al. stratified three groups according to SVR and demonstrated a lower concordance with lower SVR states.[24] This finding was reproduced by Sotomi et al. in a cardiac surgery population.[25] McLean et al. reported that in critically ill patients, after the exclusion of patients with irregular heart rhythms and aortic stenosis, that FV™ was clinically comparable to TTE in CO measurements in critically ill patients, and also made the point that multiple paired values (from the same patients) used in other studies may increase bias. No studies to date have sought to compare TTE evaluation of the HDS, or the potential clinical decision made, with FV™, or PAC.

The case for hemodynamic assessment

While rates of error may be variable with noninvasive monitoring devices, this may not affect the hemodynamic decision that is made. We chose to compare HDS rather than absolute variables for a number of reasons. Management of a patient's HDS is rarely based on a single parameter, but on an overall assessment of the cardiovascular system. While in many cases, clear parameters do not exist for treatment decisions, a combination of variables is frequently used to interpret and act on the HDS. We have taken a pragmatic view of this decision-making in tandem with proprietary suggestions regarding monitoring values, and Australasian practice. TTE is performed regularly on patients who have acute hemodynamic disturbance and algorithms exist for emergent interpretation of the HDS (e.g., RACE, H.A.R.T.scan®). A difference in variables has less clinical significance or impact if it does not alter treatment. It has been suggested that the use of PACs does not alter outcomes in large critical care populations,[3] and that the introduction of FV™ has not been associated with increased mortality.[26] Data supporting the safety of FV™ come from Kirton et al., who demonstrated no significant change in mortality over 5 years period of change from PAC to FV™ monitoring in a surgical intensive care population. Therefore, we designed the trial to assess the clinically relevant endpoint of HDS, rather than performing further measurements. Our trial examined hemodynamic data which may represent decision points in the postoperative setting (e.g., empty = fluid therapy, vasodilated = vasopressor, and PSF = inotrope).

Potential for decreased intervention

Our data is consistent with previous studies, with a poor absolute correlation between devices, and a tendency for noninvasive monitors to over-record abnormal HDSs when compared to PAC/echocardiography.[10],[14],[21],[27],[28] An important role of any monitoring device is to guide the clinician as to when no intervention is required. Monitoring devices may promote active management if staff react to abnormal parameters. This trial showed a large potential difference in active intervention between the 3 hemodynamic assessment modalities and that a less interventional style of management may be enhanced by the use of TEE/TTE.

Strengths

Our trial was a pragmatic design, and was the first to utilize algorithmic echocardiography to determine the HDS in comparison to two monitoring devices (PAC and FV™). Efforts were made to ensure a standardized system of measurement (modified H.A.R.T.scan® criteria, practice-based monitoring parameters) and standardization (all TTE/TEE practitioners had completed PTEeXAM®/H.A.R.T.scan®/PGDipCU).

All TEE/TTEs performed were independently evaluated by practitioners who were blinded to the monitoring data. “Normal” values for FV™ and PAC can be disputed; however, proprietary guidelines were used where possible, in tandem with unit practice. Several studies have shown a lack of correlation between PAC and FV™.[10],[27],[28] This trial allowed a third hemodynamic assessment modality to be applied in order to assess which device is more representative of the HDS.

We targeted the main HDSs which lead to intervention perioperatively in the cardiothoracic patient population,[29] with measures of preload, afterload, and contractility. We used simple, reproducible, echocardiographic definitions with parameters which have been studied in critical care.[16]

Limitations

This was a small, observational trial, which contained a small number of patients with abnormal HDSs. Patients with poor ejection fractions (<20%) were not evaluated as a separate patient population.

Although blinding was maintained at the time of data collection, the timing of the devices is not uniform. FV™ updates the CO approximately every 20 s while a Vigilance II™ monitor averages readings over a longer time duration, and average values can be delayed by several minutes. Limited TTE/TEE may take several minutes to perform. These circumstances are similar to the real practice of intensive care, which frequently evolves the performance of a TTE/TEE several minutes after being triggered by a PAC/FV™ reading. Measurements in the trial were taken, as far as reasonably possible, simultaneously.

While the algorithm for HDSs was constructed to mirror local practice and evidence base as closely as possible [Appendix 1] this may not reflect practice in all sites. Many sites do not consider CVP in assessment of the HDS, and there is no broad consensus as to interpretation of this variable. In addition, no assessment of right heart function was undertaken. Right heart function may influence both CVP and cardiac performance. CVP remains a commonly measured and utilized measurement in the perioperative care of the cardiac surgery patient, in a manner that is reflective in the pragmatic design of this trial.[17]

While efforts were made to standardize TTE/TEE performance and measurement, different techniques, patient habitus, and echo windows introduce variability, which is reflected in the actual practice of critical care echocardiography. A number of TTEs were technically difficult, which mirrors the cardiothoracic postoperative population.


   Conclusions Top


In this trial, FV™ and PAC were shown to agree poorly with TTE/TEE assessment of the HDS in an adult cardiothoracic population. Agreement between the FV™ and PAC was also poor. Caution is recommended in interpreting isolated hemodynamic monitoring data. All hemodynamic monitoring devices have inherent sources of error. Caution is advised in interpreting any single device or measurement as a gold standard. We suggest that hemodynamic measuring devices such as FV™/PAC may act as triggers for a global hemodynamic assessment including consideration of TTE/TEE.

Acknowledgments

H.A.R.T.scan® data collection forms were used with kind permission of Professor Colin Royse.

FV™ monitors were loaned by Edwards Lifesciences for the duration of this trial.

Financial support and sponsorship

FV™ monitors were loaned by Edwards Lifesciences for the duration of the trial however consumables were funded by the hospital. Edwards Lifesciences were not involved in trial design, or the trial itself including data analysis and write-up.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Richard C, Monnet X, Teboul JL. Pulmonary artery catheter monitoring in 2011. Curr Opin Crit Care 2011;17:296-302.  Back to cited text no. 1
    
2.
Koo KK, Sun JC, Zhou Q, Guyatt G, Cook DJ, Walter SD, et al. Pulmonary artery catheters: Evolving rates and reasons for use. Crit Care Med 2011;39:1613-8.  Back to cited text no. 2
    
3.
Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, et al. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): A randomised controlled trial. Lancet 2005;366:472-7.  Back to cited text no. 3
    
4.
Abreu AR, Campos MA, Krieger BP. Pulmonary artery rupture induced by a pulmonary artery catheter: A case report and review of the literature. J Intensive Care Med 2004;19:291-6.  Back to cited text no. 4
    
5.
Argueta E, Berdine G, Pena C, Nugent KM. FloTrac® monitoring system: What are its uses in critically ill medical patients? Am J Med Sci 2015;349:352-6.  Back to cited text no. 5
    
6.
Ganter MT, Alhashemi JA, Al-Shabasy AM, Schmid UM, Schott P, Shalabi SA, et al. Continuous cardiac output measurement by un-calibrated pulse wave analysis and pulmonary artery catheter in patients with septic shock. J Clin Monit Comput 2016;30:13-22.  Back to cited text no. 6
    
7.
Maeda T, Yoshitani K, Inatomi Y, Ohnishi Y. Inaccuracy of the FloTrac/Vigileo system in patients with low cardiac index. J Cardiothorac Vasc Anesth 2014;28:1521-6.  Back to cited text no. 7
    
8.
Suehiro K, Tanaka K, Mikawa M, Uchihara Y, Matsuyama T, Matsuura T, et al. Improved performance of the fourth-generation FloTrac/Vigileo system for tracking cardiac output changes. J Cardiothorac Vasc Anesth 2015;29:656-62.  Back to cited text no. 8
    
9.
Romagnoli S, Ricci Z, Romano SM, Dimizio F, Bonicolini E, Quattrone D, et al. FloTrac/Vigileo (third generation) and MostCare®/PRAM versus echocardiography for cardiac output estimation in vascular surgery. J Cardiothorac Vasc Anesth 2013;27:1114-21.  Back to cited text no. 9
    
10.
Desebbe O, Henaine R, Keller G, Koffel C, Garcia H, Rosamel P, et al. Ability of the third-generation FloTrac/Vigileo software to track changes in cardiac output in cardiac surgery patients: A polar plot approach. J Cardiothorac Vasc Anesth 2013;27:1122-7.  Back to cited text no. 10
    
11.
Marqué S, Gros A, Chimot L, Gacouin A, Lavoué S, Camus C, et al. Cardiac output monitoring in septic shock: Evaluation of the third-generation Flotrac-Vigileo. J Clin Monit Comput 2013;27:273-9.  Back to cited text no. 11
    
12.
Kim JY, Kim BR, Lee KH, Kim KW, Kim JH, Lee SI, et al. Comparison of cardiac output derived from FloTrac/Vigileo and impedance cardiography during major abdominal surgery. J Int Med Res 2013;41:1342-9.  Back to cited text no. 12
    
13.
McLean AS, Huang SJ, Kot M, Rajamani A, Hoyling L. Comparison of cardiac output measurements in critically ill patients: FloTrac/Vigileo vs. transthoracic Doppler echocardiography. Anaesth Intensive Care 2011;39:590-8.  Back to cited text no. 13
    
14.
Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK. Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 2007;99:329-36.  Back to cited text no. 14
    
15.
Porhomayon J, El-Solh A, Papadakos P, Nader ND. Cardiac output monitoring devices: An analytic review. Intern Emerg Med 2012;7:163-71.  Back to cited text no. 15
    
16.
Royse CF, Haji DL, Faris JG, Veltman MG, Kumar A, Royse AG. Evaluation of the interpretative skills of participants of a limited transthoracic echocardiography training course (H.A.R.T.scan course). Anaesth Intensive Care 2012;40:498-504.  Back to cited text no. 16
    
17.
Roy RC. Central venous pressure in clinical care algorithms: Are anesthesiologists and intensivists ready? Anesth Analg 2010;111:591-2.  Back to cited text no. 17
    
18.
Magder S, Potter BJ, Varennes BD, Doucette S, Fergusson D; Canadian Critical Care Trials Group. Fluids after cardiac surgery: A pilot study of the use of colloids versus crystalloids. Crit Care Med 2010;38:2117-24.  Back to cited text no. 18
    
19.
Magder S. Hemodynamic monitoring in the mechanically ventilated patient. Curr Opin Crit Care 2011;17:36-42.  Back to cited text no. 19
    
20.
Compton FD, Zukunft B, Hoffmann C, Zidek W, Schaefer JH. Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac/Vigileo) in haemodynamically unstable patients. Br J Anaesth 2008;100:451-6.  Back to cited text no. 20
    
21.
Slagt C, de Leeuw MA, Beute J, Rijnsburger E, Hoeksema M, Mulder JW, et al. Cardiac output measured by uncalibrated arterial pressure waveform analysis by recently released software version 3.02 versus thermodilution in septic shock. J Clin Monit Comput 2013;27:171-7.  Back to cited text no. 21
    
22.
Monnet X, Lahner D. Can the “FloTrac” really track flow in septic patients? Intensive Care Med 2011;37:183-5.  Back to cited text no. 22
    
23.
Concha MR, Mertz VF, Cortínez LI, González KA, Butte JM. Pulse contour analysis and transesophageal echocardiography: A comparison of measurements of cardiac output during laparoscopic colon surgery. Anesth Analg 2009;109:114-8.  Back to cited text no. 23
    
24.
Suehiro K, Tanaka K, Funao T, Matsuura T, Mori T, Nishikawa K. Systemic vascular resistance has an impact on the reliability of the Vigileo-FloTrac system in measuring cardiac output and tracking cardiac output changes. Br J Anaesth 2013;111:170-7.  Back to cited text no. 24
    
25.
Sotomi Y, Iwakura K, Higuchi Y, Abe K, Yoshida J, Masai T, et al. The impact of systemic vascular resistance on the accuracy of the FloTrac/Vigileo system in the perioperative period of cardiac surgery: A prospective observational comparison study. J Clin Monit Comput 2013;27:639-46.  Back to cited text no. 25
    
26.
Kirton OC, Calabrese RC, Staff I. Increasing use of less-invasive hemodynamic monitoring in 3 specialty surgical intensive care units: A 5-year experience at a tertiary medical center. J Intensive Care Med 2015;30:30-6.  Back to cited text no. 26
    
27.
Opdam HI, Wan L, Bellomo R. A pilot assessment of the FloTrac cardiac output monitoring system. Intensive Care Med 2007;33:344-9.  Back to cited text no. 27
    
28.
Monnet X, Anguel N, Jozwiak M, Richard C, Teboul JL. Third-generation FloTrac/Vigileo does not reliably track changes in cardiac output induced by norepinephrine in critically ill patients. Br J Anaesth 2012;108:615-22.  Back to cited text no. 28
    
29.
Guarracino F, Bertini P. Perioperative haemodynamic management: Is echocardiography the right tool? Curr Opin Crit Care 2014;20:431-7.  Back to cited text no. 29
    



 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7]


This article has been cited by
1 Consenso colombiano de cuidados perioperatorios en cirugía cardiaca del paciente adulto
José Luis Accini Mendoza,Luis Horacio Atehortua L,Victor Hugo Nieto Estrada,Carlos Eduardo Rebolledo M,Julio Cesar Duran Pérez,Juan Manuel Senior,Edgar Hernández Leiva,Albert Alexander Valencia,José Fernando Escobar Serna,Carmelo Dueñas Castell,Rafael Cotes Ramos,Nelly Beltrán,Rafael Thomen Palacio,David Andrés López García,Camilo Pizarro Gómez,María Cristina Florián Pérez,Sergio Franco,Hans García,Francisco Mauricio Rincón,Carlos Alberto Danetra Novoa,Juan Francisco Delgado
Acta Colombiana de Cuidado Intensivo. 2020;
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Methods
   Results
   Discussion
   Conclusions
    References
    Article Tables

 Article Access Statistics
    Viewed2550    
    Printed59    
    Emailed0    
    PDF Downloaded167    
    Comments [Add]    
    Cited by others 1    

Recommend this journal